Call Us Toll Free: 1-800-458-6153
Return To Main Interscan Home Page  |  Products  |  Tech Center  |  E-mail Us
Search Catalog
      GO   
 All Categories    Single Point Stationary Monitors - LD Series - Hydrogen Peroxide
All Categories > Gas Detection Instruments And Systems > Single and 2-Point Stationary Area Monitors > Single Point Stationary Monitors - LD Series - Hydrogen Peroxide > Item # LD09-1000m  

Item # LD09-1000m, Single Point Stationary Monitors - LD Series - Hydrogen Peroxide


    border="0" Printable Page  
    Email Email This Page  
    Save To Favorites Save To Favorites
  

Single Point LD Series Analyzers include:
  • 3-1/2 digit LED display
  • Appropriate sensor
  • Integral sample pump and rotameter
  • Single-Point analyzer housed in NEMA 4X fiberglass enclosure with window.
  • Millipore particulate filter
  • Dual alarm set points activating relay contact closures
  • 4-20 mA and 0 - 1 V output
  • Power cord
  • Sensor protection circuit (for EtO monitors only)
  • Interference scrubber (for CO, NO, & EtO monitors only)






Specifications

Accuracy

± 2.0% of reading, ± 1 least significant digit. (Limited to accuracy of calibration standard)

Repeatability

± 0.5% of full scale

Minimum Detectability

1.0% of full scale

Linearity

± 1.0% of full scale

Zero Drift

± 1.0% of full scale (24 hours)

Span Drift

Less than ± 2.0% of full scale (24 hours)

Calibration

Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service.

Lag Time

Less than 1 second

Range

0-1000 ppm

Resolution

1 ppm

Dimensions

22" H x 18" W x 8.5" D    (559 x 457 x 216 mm)

Weight

22 lb   (10.0 kg)

Gas

Hydrogen Peroxide (H2O2)

Number of Detection Points

1

Application

Continuous

Includes Auto Zero

No

Note

The zero and span drift specifications assume that the analyzer is equilibrated, and is at constant temperature, with a properly maintained sensor.


 DOWNLOADS 

Gas Detection Applications Primer

Sensor Principle of Operation

Interfering Gas Data

Gas Detection Knowledge Base

LD Series (Single Point) Instruction Manual
(PDF, 0KB)

Hydrogen Peroxide Monitoring Applications

How Do I Get Started?

Interscan Photo Gallery

Interscan Video Library

H2O2-Toxic and Hazardous Properties




© Interscan Corporation
4590 Ish Drive #110 • Simi Valley, CA 93063–7682
TOLL FREE: (800) 458-6153 (US & Canada) • PHONE: (818) 882-2331
FAX: (818) 341-0642 • E-MAIL: info@gasdetection.com