Call Us Toll Free: 1-800-458-6153
Return To Main Interscan Home Page  |  Products  |  Tech Center  |  E-mail Us
Search Catalog
      GO   
 All Categories    Portable Analyzer (4000 Series) with Digital Display - Peracetic acid
All Categories > Gas Detection Instruments And Systems > Portable Analyzers > Portable Analyzer (4000 Series) with Digital Display - Peracetic acid > View Items  

Portable Analyzer (4000 Series) with Digital Display - Peracetic acid

Check up to five results to perform an action.

4000 Series Portable Analyzer Includes:
  • 3-1/2 digit LCD display
  • Appropriate sensor
  • Integral sample pump & plug-in charger
  • User-adjustable single alarm level activating audible and visual alarm
  • 0-100 mV output
  • Carrying case with shoulder strap
  • Interference scrubber
  • 12 inch (30.5 cm) Teflon sample probe


Search Items



    border="0" Printable Page  
    Email Email This Page  
    Save To Favorites Save To Favorites
  
Compare Items
   
  Results 1 - 3 of 3 1 

Item #

Accuracy

Repeatability

Minimum Detectability

Linearity

Zero Drift

Span Drift

Calibration

Lag Time

Rise time to 90% of final value (sec)

Rise time to 50% of final value (sec)

Fall time to 10% of original value (sec)

Range

Resolution

Dimensions

Weight

4890-5.00m ± 2.0% of reading, ± 1 least significant digit. (Limited to accuracy of calibration standard) ± 0.5% of full scale 1.0% of full scale ± 1.0% of full scale ± 1.0% of full scale (24 hours) Less than ± 2.0% of full scale (24 hours) Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service. Less than 1 second 45 10 30-45 0-5.00 ppm 0.01 ppm 7" H x 4" W x 8.875" D   (178 x 102 x 225 mm) 4.5 lb    (2.0 kg)
4890-19.99m ± 2.0% of reading, ± 1 least significant digit. (Limited to accuracy of calibration standard) ± 0.5% of full scale 1.0% of full scale ± 1.0% of full scale ± 1.0% of full scale (24 hours) Less than ± 2.0% of full scale (24 hours) Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service. Less than 1 second 45 10 30-45 0-19.99 ppm 0.01 ppm 7" H x 4" W x 8.875" D   (178 x 102 x 225 mm) 4.5 lb    (2.0 kg)
4890-50.0m ± 2.0% of reading, ± 1 least significant digit. (Limited to accuracy of calibration standard) ± 0.5% of full scale 1.0% of full scale ± 1.0% of full scale ± 1.0% of full scale (24 hours) Less than ± 2.0% of full scale (24 hours) Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service. Less than 1 second 45 10 30-45 0-50.0 ppm 0.1 ppm 7" H x 4" W x 8.875" D   (178 x 102 x 225 mm) 4.5 lb    (2.0 kg)
  Results 1 - 3 of 3 1 


© Interscan Corporation
4590 Ish Drive #110 • Simi Valley, CA 93063–7682
TOLL FREE: (800) 458-6153 (US & Canada) • PHONE: (818) 882-2331
FAX: (818) 341-0642 • E-MAIL: info@gasdetection.com